Total hip replacement is one of the most common procedures in orthopaedic surgery: around 165,000 are performed each year in France, nearly a third of them after a femoral neck fracture [1]. The operation is well established, but its outcome depends largely on the position of the implants.
3D planning addresses precisely this point: defining in advance, on the patient’s digital twin, the strategy to reproduce in theatre. It is increasingly used for complex anatomies. Custom surgical guides are also appearing, as one way of transferring this plan to the operating theatre, but their use is still at an early stage. This article details the workflow, from CT scan to device, what the studies say and the regulatory framework.
It is aimed at orthopaedic surgeons, biomedical engineers and medical device manufacturers. External appliances (orthoses, limb prostheses) follow a different chain, presented on our orthotists and prosthetists page.
Why plan a hip replacement in 3D?
3D planning makes it possible to set, before the operation, the parameters that determine the stability and longevity of the prosthesis. On a pelvic X-ray, the surgeon works on a projection: magnification, limb rotation and acetabular anteversion are poorly assessed.
The CT scan gives access to the patient’s actual anatomy, in volume. Four objectives guide the planning:
- restoring the centre of rotation of the hip, often displaced by wear or dysplasia;
- equalising limb length and restoring the femoral offset, which determines the tension of the gluteal muscles;
- orienting the cup in inclination and anteversion, to limit the risk of dislocation and wear;
- choosing the implant size before opening the boxes in theatre.
Lower limb length is measured and corrected before the operation, rather than adjusted “by eye” in theatre: this is one of the concrete benefits of planning.
2D or 3D planning: what are the differences?
2D planning on X-ray remains the reference for a standard THR; 3D becomes useful as soon as the anatomy departs from the norm [6]. The table below summarises the differences between the two approaches.
Faites défiler le tableau horizontalement →
| Critère | Planification 2D (radiographie) | Planification 3D (scanner) |
|---|---|---|
| Imagerie de départ | Radiographie du bassin de face, avec marqueur de calibration | Scanner du bassin et du fémur, au format DICOM |
| Mesures | Projections, sensibles à l'agrandissement et à la rotation du membre | Mesures en volume, indépendantes de la position du patient |
| Antéversion du cotyle et version du col fémoral | Estimée ou non mesurable | Mesurée directement |
| Simulation | Calques d'implants superposés à l'image | Implants positionnés sur le jumeau numérique, simulation de la mobilité et des conflits |
| Dispositifs dérivés | Aucun | Modèles anatomiques, implants sur mesure, et, de façon encore émergente, guides chirurgicaux |
| Contraintes | Rapide, peu coûteuse, faible irradiation | Coût et dose d'irradiation plus élevés, segmentation à réaliser |
A CT scan is not routine for a standard THR: it is mainly justified when the anatomy is complex, at the price of higher cost and radiation exposure.
From CT scan to surgical guide: the six-step workflow
When a custom guide is chosen, it forms part of a six-step digital chain, from imaging to post-operative checks. Each step produces a file that feeds the next. The first three already serve planning alone, with or without a guide.

- Image acquisition. A CT scan of the pelvis, often supplemented by an acquisition of the knee to measure femoral version, is performed a few weeks before the operation. Thin slices limit artefacts during reconstruction.
- Segmentation. The DICOM data are imported into medical imaging segmentation software such as Simpleware. The pelvis and femur are isolated, manually or through AI-assisted segmentation, to obtain the patient’s digital twin.
- Planning. The team virtually positions the cup and stem, checks the centre of rotation, offset and length, then simulates the range of motion. The surgeon validates the plan.
- Guide design. If a guide is planned, it is designed in CAD from the bone surface: one side fits the rim of the acetabulum or the femoral neck, the other carries a drill sleeve or a cutting slot. Organic modelling software such as Geomagic Freeform, operated with the Touch haptic arm or a mouse, makes this work easier on irregular anatomical shapes.
- Manufacturing and sterilisation. The guide and, if needed, a 3D anatomical model of the pelvis are printed in a biocompatible, sterilisable material, by the manufacturer or a hospital platform. The model makes it possible to test the guide’s fit before the operation.
- Use in theatre and checks. The surgical guide is placed on its unique contact area; it directs a reference pin or the cutting / drilling plane. A post-operative CT scan can measure the difference between the plan and the position achieved.
Steps 2 to 4 are the most time-consuming and the most repetitive from one case to the next. Automation plugins such as Dynabot, developed on Geomagic Freeform, can standardise guide design for a given protocol.
Which custom devices for hip replacement?
3D planning can lead to five families of devices, from the disposable instrument to the permanent implant. Patient-specific 3D implants are reserved for major reconstructions. 3D surgical guides, for their part, remain an emerging use in hip surgery: they are mainly used by pioneering teams and in difficult cases [13] [14].
Faites défiler le tableau horizontalement →
| Dispositif | Rôle au bloc | Situation typique | Matériau courant |
|---|---|---|---|
| Guide d'orientation cotyloïdien | Positionne une broche qui donne l'axe d'impaction de la cupule | Cotyle dysplasique ou déformé ; usage émergent, y compris en première intention dans quelques équipes | Polymère stérilisable (PA12, résine biocompatible) |
| Guide de coupe du col fémoral | Fixe le niveau et l'orientation de l'ostéotomie du col | Contrôle de la longueur et de la version de la tige ; usage émergent | Polymère stérilisable |
| Guide de perçage | Oriente les vis de fixation d'un implant acétabulaire | Reprise avec implant sur mesure | Polymère stérilisable |
| Modèle anatomique du bassin | Visualiser la perte osseuse, tester le guide et l'implant avant l'intervention | Reprise, dysplasie, tumeur | Résine ou polymère technique |
| Implant acétabulaire sur mesure (cupule « triflange ») | Comble le défect et prend appui sur l'ilion, l'ischion et le pubis | Défects acétabulaires sévères (Paprosky 3A et 3B), discontinuité pelvienne [8] | Titane poreux imprimé en 3D |
When used together, the guide and the implant are often designed in a coordinated way: the drilling guide reproduces the screw positions planned on the implant, which ensures that the part sits exactly where it was planned.
What the studies say
The custom guide improves placement accuracy compared with the freehand technique, but its clinical benefit has not yet been demonstrated [3] [6]. The main published data are summarised below.
A practice that is still emerging. In 2018, a review from University College London (UCL) pointed out that these guides, well established in knee surgery, were used in the hip by only a small number of surgeons. It reported early results of accurate positioning, accompanied by higher costs and an uncertain clinical effect [13]. A British centre for complex hip surgery presents a use case [14], but they do not constitute standard practice.
Greater accuracy than standard instrumentation
- 2022 meta-analysis. Constantinescu et al. compare custom guides with standard instrumentation in primary THR [3]. The mean cup orientation error falls by 6.25° in anteversion (heterogeneity I² = 98%) and by 4.00° in inclination (I² = 85%). These differences between studies are very high: the mean figure indicates a trend, not a gain to be expected in every department. The relative risk of a cup outside the Lewinnek zone is 0.23 in anteversion and 0.12 in inclination.
- Prospective series of 100 patients. Spencer-Gardner et al. (Bone & Joint Journal, 2016) measure on post-operative CT a mean deviation from the plan of 3.9° in inclination and 3.6° in anteversion [4]. Both targets are achieved within ± 10° in 91% of cases, a level the authors consider comparable to navigation and robotics.
- Randomised trial of 54 patients. Thomas et al. (Arthroplasty, 2022) find no difference in mean anteversion error [5]. However, the anteversion achieved stays within 10° of the plan in 96% of cases with CT planning and a guide, versus 76% with the conventional technique: it is mainly the large deviations that disappear.
Indications mainly in complex cases
A review published in 2023 in the Revue médicale suisse concludes that custom guides cannot be recommended for routine hip or knee replacement [6]. They have a place in tumour surgery, acetabular architectural defects and malunions, despite a high additional cost.
The situations where the literature is most favourable:
- Severe hip dysplasia (Crowe III and IV). A randomised trial of 104 adults concludes that the guide improves cup positioning and functional recovery in Crowe III and IV dysplasia, with little difference for Crowe I and II forms [7].
- Revisions with major bone loss. A 2023 meta-analysis of 1,218 patients evaluates custom triflange cups for Paprosky 3A and 3B defects: 15% reoperations and 12% implant failures at a mean follow-up of 47 months [8]. These implants are not guides, but they illustrate the place of customisation in the most difficult cases.
“Custom-made” guide: what the regulations say
A guide printed for one patient is not necessarily a “custom-made” medical device in the regulatory sense. The European MDR (EU 2017/745) distinguishes several regimes, clarified by the MDCG 2021-3 guidance published in March 2021 [10] [11].
- Custom-made device: manufactured on the written prescription of an authorised professional, who specifies its design characteristics, for a single patient. It is exempt from CE marking but follows Annex XIII of the MDR [10]; in France, its manufacturer registers with the ANSM [12].
- “Patient-matched” device: adapted to the patient’s anatomy from a standard model or a design envelope, through a validated, reproducible industrial process. It follows the normal CE marking route. The MDCG cites knee arthroplasty cutting guides among its examples [11].
- Device manufactured and used within a healthcare institution (“in-house”): a hospital that produces guides itself for its own patients may benefit from a partial exemption, under the specific conditions set out in Article 5(5) of the MDR (quality management system, justification that no equivalent is available on the market, public declaration, documentation kept available for the competent authority) [10].
A guide mass-produced by a manufacturer from its own planning platform therefore generally falls under the second regime. In-house production in a hospital follows the third, with obligations of its own.
In either case, have the classification of your devices validated by your regulatory department or a specialist consultant before starting production, whether in series or in-house. Manufacturers in the sector will also find guidance on our page dedicated to medical device manufacturers.
Bringing 3D planning in-house with Twin 3D
Twin 3D equips orthopaedic surgery departments, hospital 3D platforms and medical device manufacturers so that they can carry out the segmentation, planning and design steps themselves. Manufacturing remains with your own equipment or our industrial partners.
- Segmentation: Simpleware reconstructs the pelvis and femur from the CT scan, with AI-assisted segmentation tools. Processing is done locally, without sending patient data to the cloud, in a few minutes.
- Design of guides and models: Geomagic Freeform, which can be combined with the Touch or Touch X haptic arm, makes it possible to sculpt contact surfaces on irregular bone shapes.
- Automation: we develop Dynabot plugins based on your own protocols, so that every hip guide follows the same steps and the same checks.
- Training: our Simpleware and Geomagic Freeform training courses take place on site or remotely; the Geomagic Freeform course concludes with an application project, for example a surgical device.
Every project begins with a proof of concept on your own data: an anonymised hip case, segmented and modelled with your teams. You can thus assess the tool on a real case before investing. Contact us to arrange one.
FAQ
A CT scan of the pelvis, exported in DICOM format, is the basis. It is often supplemented by a few slices of the knee to measure femoral version. Thin slices improve segmentation quality. MRI imaging can also be used by our software.
It depends on the manufacturing method. A guide produced on a hospital platform can be ready in a few days [9]; a custom acetabular implant may require several weeks of design, validation and manufacturing.
In a series of 100 patients, the accuracy of the guide is considered comparable to that of navigation and robotics [4]. The robot, however, requires a heavy investment and a dedicated organisation, whereas using the surgical guide requires no equipment in theatre and relies on the planning carried out beforehand. The choice depends on the volume of complex cases and the department’s resources.
Not always. According to the MDCG 2021-3 guidance [11], a guide produced by a validated industrial process from a standard model is a “patient-matched” device subject to CE marking. The custom-made regime requires a written prescription setting out the design characteristics.
Yes. A department equipped with segmentation software, a CAD tool and a suitable printer can design its own guides and anatomical models. It must at the same time comply with the MDR requirements applicable to healthcare institutions (Article 5(5)) [10], and seek advice from its regulatory department or a specialist consultant. The department may also choose to bring the segmentation and design stages in-house and delegate the printing of the guides to an external provider.
Sources
- La Revue du Praticien. Prothèse totale de la hanche : une intervention maîtrisée mais des complications possibles justifiant un suivi à long terme.
- Abdel MP et al. What safe zone? The vast majority of dislocated THAs are within the Lewinnek safe zone for acetabular component position. Clin Orthop Relat Res. 2016;474:386-391.
- Constantinescu DS et al. The efficacy of patient specific instrumentation (PSI) in total hip arthroplasty (THA): a systematic review and meta-analysis. J Orthop. 2022;34:404-413.
- Spencer-Gardner L et al. Patient-specific instrumentation improves the accuracy of acetabular component placement in total hip arthroplasty. Bone Joint J. 2016;98-B(10):1342-1346.
- Thomas C et al. Does functional planning, 3D templating and patient-specific instrumentation improve accuracy in total hip replacement? A randomized controlled trial. Arthroplasty. 2022;4:43.
- Fell H et al. Guides de coupes 3D sur mesure : quelles applications en chirurgie orthopédique ? Rev Med Suisse. 2023;19(854):2344-2349.
- Wang C et al. Patient-specific total hip arthroplasty is superior to conventional methods for Crowe III and IV adult developmental hip dysplasia: a randomized controlled trial. Ann Transl Med. 2021;9(3).
- Broekhuis D et al. Custom triflange acetabular components for large acetabular defect reconstruction in revision total hip arthroplasty: a systematic review and meta-analysis on 1218 patients. EFORT Open Rev. 2023;8(7):522-531.
- Pham Trung H et al. Early outcomes of total hip arthroplasty using point-of-care manufactured patient-specific instruments: a single university hospital's initial experience. BMC Surg. 2023;23:369.
- Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices. EUR-Lex.
- Medical Device Coordination Group. MDCG 2021-3: Questions and Answers on Custom-Made Devices & considerations on Adaptable medical devices and Patient-matched medical devices. European Commission, March 2021.
- ANSM. Custom-made medical devices.
- Henckel J et al. 3D-printed patient-specific guides for hip arthroplasty. J Am Acad Orthop Surg. 2018;26(16):e342-e348.
- Complex Hip Surgery (Prof. Alister Hart, Royal National Orthopaedic Hospital and UCL). Using 3D-printed guides to achieve the best possible position of hip replacements. Clinical case, no quantitative data.